中文
English
Home    Paper package tester    Diapers Permeability Tester ZNK-100

Diapers Permeability Tester ZNK-100

Model:ZNK-100
Brand:Sumspring

 

Diapers permeability tester ZNK-100 is used for test permeability of diapers, sanitary napkins, sanitary pads and other products.Diaper permeability tester  adjustable angle 0~45 °, to meet the different requirements of the standard angle for penetration, widely used in health products factory, quality inspection and other institutions.

Diapers Permeability Tester ZNK-100

 

Test principle

Take a sample to be tested, to expand it naturally state down on a large enough known mass filter paper and placed in the bottom of the funnel Diaper permeability tester to make sure the liquid funnel fitted quickly open the valve. Remove the sample after liquid freedom flowing the sample surface. Weigh the sample at the bottom of the known quality filter paper, expressed osmotic quantity by weight difference.

 

Application

Product

Name

Application

IMG_257

Diapers

For the permeability determination of diapers, adult diapers and other products 

IMG_258  Sanitary napkins

For the permeability test of products of sanitary napkins, sanitary pads etc..

Non vowen,Nursing pad For the permeability test of products of non vowen, nursing pads etc..

 

Technical Feature

★Diapers permeability tester look beautiful and fully equipped, user-friendly

★A variety of options to meet different standards

★Stainless steel countertops, beautiful rust-proof, simple structure, easy to operate.

★0~45 ° adjustable angle test, meet diapers, sanitary napkins, sanitary pads and other different test requirements for the angles

★Large stainless steel table is about 290 * 400mm,suitable for all kind of sanitary napkin and nursing mat.

 

Main Specifications

TECHNICAL SPECIFICATIONS

Briquetting

φ100mm,Weight(1.2±0.002)kg(Pressure 1.5kPa)

Standard funnel

A: Volume 80mL,For baby diapers / tablet / pad test
B: Volume 150mL,For adult diapers / tablet / pad test
C: Volume 60ml, For sanitary napkins test(Optional)

Experiments angle 

30°±2°(Diapers)
10°±1°(Sanitary napkins, Optional)

25°±1° non vowen

Discharge funnel  

150ml

Dimension

410mm×310mm×640mm

Net weight

18kg

ENVIRONMENT REQUIREMENT

Temperature 

15℃-35℃

Relative Humidity

No more than 70%,no dewdrop

 

Standard
GB/T28004-2011,GB/T8939-2008

 

Configuration

Diapers permeability tester host, filter paper, beakers, stainless steel standard briquetting, funnel, cylinder

 

Contact US

Note: Sumspring is committed to providing high-quality packaging testing instruments for you, some products in the process of upgrading, product parameters and appearance will change accordingly, please call the Jinan Sumspring Experiment Instrument Co. Ltd. to inquiry.

 